![Ronde & Schwarz 1304.0004.03 Скачать руководство пользователя страница 802](http://html1.mh-extra.com/html/ronde-and-schwarz/1304-0004-03/1304-0004-03_user-manual_1484985802.webp)
Remote Commands
R&S
®
FPL1000
802
User Manual 1178.3370.02 ─ 09
●
SOURce<si>:POWer[:LEVel][:IMMediate][:AMPLitude]
●
SOURce<si>:POWer[:LEVel][:IMMediate]:OFFSet
Remote commands exclusive to tracking generators:
SOURce<si>:FREQuency:COUPling[:STATe]
....................................................................802
SOURce<si>:INTernal:FREQuency:OFFSet
......................................................................802
.........................................................................................802
[SENSe:]CORRection:COLLect[:ACQuire]
........................................................................803
....................................................................................... 804
..........................................................................................804
[SENSe:]CORRection:TRANsducer:GENerate
.................................................................. 804
DISPlay[:WINDow<n>][:SUBWindow<w>]:TRACe<t>:Y[:SCALe]:RVALue
............................805
SOURce<si>:FREQuency:COUPling[:STATe]
<State>
Enables or disables the usage of the internal generator as a tracking generator. The
tracking generator is only available for frequency sweeps in the Spectrum application.
Suffix:
<si>
.
1..n
Parameters:
<State>
ON | OFF | 0 | 1
OFF | 0
Switches the tracking generator off
ON | 1
Switches the tracking generator on
*RST:
0
Example:
SOUR:FREQ:COUP:STAT ON
Manual operation:
See
SOURce<si>:INTernal:FREQuency:OFFSet
<FreqOffset>
Defines an offset of the tracking generator signal to the analyzer frequency.
Suffix:
<si>
.
1..n
Parameters:
<FreqOffset>
Default unit: HZ
Example:
SOUR:INT:FREQ:OFFS 10kHz
Manual operation:
See
[SENSe:]CORRection[:STATe]
<State>
This command turns correction of measurement results (normalization) on and off.
Setting Basic Spectrum RF Measurement Parameters