SYSTem:FIRMware:UPDate
<Directory>
This command starts a firmware update using the *.msi files in the selected directory.
The default path is
D:\FW_UPDATE
command. To store the update files the
Only user accounts with administrator rights can perform a firmware update.
Setting parameters:
<Directory>
Example:
SYST:FIRM:UPD 'D:\FW_UPDATE'
Starts the firmware update from directory "
D:\FW_UPDATE
".
SYSTem:FORMat:IDENt
<IDNFormat>
This command selects the response format to the
Parameters:
<IDNFormat>
NEW | FSL
R&S
FPL1000 format
Format is also compatible to the R&S FSL and R&S FSV family
*RST:
not reset!
11.10.8
Using Service Functions
......................................................................................975
DIAGnostic:SERVice:SFUNction:LASTresult?
................................................................... 976
DIAGnostic:SERVice:SFUNction:RESults:DELete
............................................................. 976
DIAGnostic:SERVice:SFUNction:RESults:SAVE
................................................................976
.......................................................................................... 976
........................................................................................977
............................................................................................. 977
DIAGnostic:SERVice:SFUNction
<ServiceFunction>
This command starts a service function.
The service functions are available after you have entered the level 1 or level 2 system
password.
Parameters:
<ServiceFunction>
String containing the ID of the service function.
The ID of the service function is made up out of five numbers,
separated by a point.
•
function group number
•
board number
•
function number
•
parameter 1 (see the Service Manual)
•
parameter 2 (see the Service Manual)
Configuring the R&S