Remote Commands
R&S
®
FPL1000
786
User Manual 1178.3370.02 ─ 09
Manual operation:
See
See
See
See
See
See
TRIGger[:SEQuence]:TIME:RINTerval
<Interval>
This command defines the repetition interval for the time trigger.
Parameters:
<Interval>
2.0 ms to 5000
Range:
2 ms to 5000 s
*RST:
1.0 s
Default unit: S
Example:
TRIG:SOUR TIME
Selects the time trigger input for triggering.
TRIG:TIME:RINT 50
The sweep starts every 50 s.
Manual operation:
See
11.8.4.2
Configuring Gated Measurements
....................................................................................787
.....................................................................................787
....................................................................................787
................................................................................... 788
........................................................................................788
[SENSe:]SWEep:EGATe
<State>
This command turns gated measurements on and off.
The measurement ends when a particular number of measurement points has been
recorded.
(See
[SENSe:]SWEep[:WINDow<n>]:POINts
Performing gated measurements turns the squelch off.
Parameters:
<State>
ON | OFF | 0 | 1
OFF | 0
Switches the function off
ON | 1
Switches the function on
Setting Basic Spectrum RF Measurement Parameters