![Ronde & Schwarz 1304.0004.03 Скачать руководство пользователя страница 347](http://html1.mh-extra.com/html/ronde-and-schwarz/1304-0004-03/1304-0004-03_user-manual_1484985347.webp)
The Spectrum Application (RF Measurements)
R&S
®
FPL1000
347
User Manual 1178.3370.02 ─ 09
●
Softkeys for active markers (displayed on the screen) are highlighted blue. The
softkey for the currently selected marker (for which functions are performed) is
highlighted orange.
●
To set individual markers very quickly, use the softkeys in the "Marker" menu.
●
To set up several markers at once, use the "Marker" dialog box.
●
To position the selected marker to a special value, use the softkeys in the "Marker
To" menu.
●
To determine more sophisticated marker results, use the special functions in the
"Marker Function" dialog box.
In addition to basic markers, sophisticated marker functions are provided for special
results such as noise or band power measurements.
.................................................................................................347
7.9.1.1
Marker Types
All markers can be used either as normal markers or delta markers. A normal marker
indicates the absolute signal value at the defined position in the diagram. A delta
marker indicates the value of the marker relative to the specified reference marker (by
default marker 1).
In addition, special functions can be assigned to the individual markers. The availability
of special marker functions depends on whether the measurement is performed in the
frequency or time domain, and on the type of measurement.
Temporary markers are used in addition to the markers and delta markers to analyze
the measurement results for special marker functions. They disappear when the asso-
ciated function is deactivated.
7.9.1.2
Activating Markers
Only active markers are displayed in the diagram and in the marker table.
Active markers are indicated by a highlighted softkey.
By default, marker 1 is active and positioned on the maximum value (peak) of trace 1
as a normal marker. If several traces are displayed, the marker is set to the maximum
value of the trace which has the lowest number and is not frozen (View mode). The
next marker to be activated is set to the frequency of the next lower level (next peak)
as a delta marker; its value is indicated as an offset to marker 1.
A marker can only be activated when at least one trace in the corresponding window is
visible. If a trace is switched off, the corresponding markers and marker functions are
also deactivated. If the trace is switched on again, the markers along with coupled
functions are restored to their original positions, provided the markers have not been
used on another trace.
Marker Usage