Remote Commands
R&S
®
FPL1000
688
User Manual 1178.3370.02 ─ 09
//Define a relative limit of -20 dBc for the entire range 2 for power class 2.
ESP:RANG2:LIM2:REL:STAR -20
ESP:RANG2:LIM2:REL:STOP -20
//------------ Configuring List Evaluation-----------------
//Activate list evaluation, i.e. the peak is determined for each range
//after each sweep.
CALC:ESP:PSE:AUTO ON
//Define a peak threshold of 10 dB.
CALC:ESP:PSE:MARG 10dB
//------------ Managing Measurement Configurations----------
//Save the current configuration in a new file named '3GPP_UL_User'
//in the same directory so the standard is not overwritten.
ESP:PRES:STOR 'WCDMA\3GPP\UL\3GPP_UL_User.xml'
//------------ Performing the measurement-------------------
//One sweep
INIT:ESP
//------------ Checking the Results-------------------------
//Query the result of the limit check for all ranges.
CALC:LIM:FAIL?
//Query the peak for each range of the SEM measurement as a list.
TRAC:DATA? LIST
11.6.8
Measuring Spurious Emissions
The following commands are required to perform spurious emissions measurements.
................................................................................. 688
...................................................................................... 689
Configuring the List Evaluation
............................................................................. 697
Adjusting the X-Axis to the Range Definitions
.......................................................699
Performing a Spurious Measurement
................................................................... 699
Retrieving and Saving Settings and Results
.........................................................699
Programming Example: Spurious Emissions Measurement
.................................699
11.6.8.1
Initializing the Measurement
Note that with the R&S
FPL1000, the spurious measurement must be initialized before
you can start configuring the sweep list or list evaluation.
INITiate<n>:SPURious
This command initiates a Spurious Emission measurement.
Configuring and Performing Measurements