Getting Started
R&S
®
FPL1000
38
User Manual 1178.3370.02 ─ 09
Pin
Description
1
2
3
1
Plus
2
Ground
3
Not used
If you use an external power supply unit to supply safety extra-low DC voltage (SELV)
to the instrument, be sure to meet the requirements for reinforced/double insulation in
accordance with DIN/EN/IEC 61010 (UL 3111, CSA C22.2 No. 1010.1) or DIN/EN/IEC
60950 (UL 1950, CSA C22.2 No. 950). Provide current limitation in accordance with
DIN EN 61010-1 Appendix F2.1.
Also see
5.2.2.2
AC Power Supply Connection and Main Power Switch
An AC power supply connector and main power switch are located in a unit on the rear
panel of the instrument.
Main power switch function:
Position 1: The instrument is in operation.
Position O: The entire instrument is disconnected from the AC power supply.
5.2.2.3
GPIB Interface
The optional GPIB interface (R&S
FPL1-B10) is in compliance with IEEE488 and
SCPI. A computer for remote control can be connected via this interface. To set up the
connection, a shielded cable is recommended.
For details, see
Chapter 10.1.1.2, "GPIB Interface (IEC 625/IEEE 418 Bus Interface)"
5.2.2.4
Ref. In / Ref. Out
The Ref. In connectors are used to provide an external reference signal to the
R&S
FPL1000.
The Ref. Out connectors can be used to provide a reference signal from the
R&S
FPL1000 to other devices that are connected to this instrument.
Various connectors are provided for different reference signals: