Remote Commands
R&S
®
FPL1000
772
User Manual 1178.3370.02 ─ 09
[SENSe:]BWIDth:VIDeo:TYPE
<Mode>
[SENSe:]BANDwidth:VIDeo:TYPE
<Mode>
This command selects the position of the video filter in the signal path.
Parameters:
<Mode>
LINear
The video filter is applied in front of the logarithmic amplifier.
In linear mode, measurements with a logarithmic level scale
result in flatter falling edges compared to logarithmic mode. The
reason is the conversion of linear power values into logarithmic
level values: if you halve the linear power, the logarithmic level
decreases by 3
dB.
LOGarithmic
The video filter is applied after the logarithmic amplifier
*RST:
LOGarithmic
Example:
BAND:VID:TYPE LIN
Video filter ahead of the logarithmic amplifier
11.8.2.2
Configuring the Sweep
Useful commands for configuring sweeps described elsewhere:
●
●
●
Remote commands exclusive to configuring sweeps:
.................................................................................................772
........................................................................................... 773
.............................................................................................773
[SENSe:]SWEep[:WINDow<n>]:POINts
............................................................................774
......................................................................................... 775
.........................................................................................775
[SENSe:]SWEep:COUNt
<SweepCount>
This command defines the number of sweeps that the application uses to average
traces.
In continuous sweep mode, the application calculates the moving average over the
average count.
In single sweep mode, the application stops the measurement and calculates the aver-
age after the average count has been reached.
Setting Basic Spectrum RF Measurement Parameters