![Ronde & Schwarz 1304.0004.03 Скачать руководство пользователя страница 277](http://html1.mh-extra.com/html/ronde-and-schwarz/1304-0004-03/1304-0004-03_user-manual_1484985277.webp)
The Spectrum Application (RF Measurements)
R&S
®
FPL1000
277
User Manual 1178.3370.02 ─ 09
"Auto"
The R&S
FPL1000 determines which IF path to use automatically,
depending on the used analysis bandwidth.
"Off"
The wide IF path is always used.
Remote command:
7.3.2
Power Sensors
The R&S
FPL1000 can also analyze data from a connected power sensor.
The "Sensor" connector is provided by the "Additional Interfaces" option R&S
FPL1-
B5. Additionally, the power sensor measurement requires the option R&S
FPL1-K9.
..................................................................................... 277
..........................................................................................278
How to Work With a Power Sensor
.......................................................................281
7.3.2.1
Basics on Power Sensors
For precise power measurement, up to 4 power sensors can be connected to the
instrument via the optional power sensor interface (on the rear panel) or the USB con-
nectors. Both manual operation and remote control are supported.
For a detailed list of supported sensors, see the data sheet.
Signal
source
Power
sensor
Signal
analyzer
Figure 7-41: Power sensor support – standard test setup
Using the power sensor with several applications
The power sensor cannot be used from the R&S
FPL1000 firmware and the R&S
Power Viewer Plus (virtual power meter for displaying results of the R&S NRP power
sensors) simultaneously.
Result display
The results of the power sensor measurements are displayed in the marker table. For
each power sensor, a row is inserted. The sensor index is indicated in the "Type" col-
umn.
Receiving Data Input and Providing Data Output