![Ronde & Schwarz 1304.0004.03 Скачать руководство пользователя страница 712](http://html1.mh-extra.com/html/ronde-and-schwarz/1304-0004-03/1304-0004-03_user-manual_1484985712.webp)
Remote Commands
R&S
®
FPL1000
712
User Manual 1178.3370.02 ─ 09
The number of results required for the calculation of average is defined with
.
Averaging is reset by switching it off and on again.
Synchronization to the end of averaging is only possible in single sweep mode.
Suffix:
<n>
.
<m>
Parameters:
<State>
ON | OFF | 1 | 0
*RST:
0
Example:
INIT:CONT OFF
Switches to single sweep mode.
CALC:MARK:FUNC:SUMM:AVER ON
Switches on the calculation of average.
AVER:COUN 200
Sets the measurement counter to 200.
INIT;*WAI
Starts a sweep and waits for the end.
CALCulate<n>:MARKer<m>:FUNCtion:SUMMary:PHOLd
<State>
This command switches on or off the peak-hold function for the active power measure-
ment in zero span in the window specified by the suffix <n>. If activated, the peak for
each sweep is compared to the previously stored peak; the maximum of the two is
stored as the current peak.
The peak-hold function is reset by switching it off and on again.
Suffix:
<n>
.
<m>
Parameters:
<State>
ON | OFF | 1 | 0
*RST:
0
CALCulate<n>:MARKer<m>:FUNCtion:SUMMary[:STATe]
<State>
This command turns time domain power measurements on and off. This measurement
is only available in zero span.
When you turn the measurement on, the R&S
FPL1000 activates a marker and posi-
tions it on the peak power level in the marker search range.
Suffix:
<n>
.
<m>
Configuring and Performing Measurements