![Ronde & Schwarz 1304.0004.03 Скачать руководство пользователя страница 846](http://html1.mh-extra.com/html/ronde-and-schwarz/1304-0004-03/1304-0004-03_user-manual_1484985846.webp)
Remote Commands
R&S
®
FPL1000
846
User Manual 1178.3370.02 ─ 09
DISPlay[:WINDow<n>]:MTABle
<DisplayMode>
This command turns the marker table on and off.
Suffix:
<n>
.
irrelevant
Parameters:
<DisplayMode>
ON | 1
Turns on the marker table.
OFF | 0
Turns off the marker table.
AUTO
Turns on the marker table if 3 or more markers are active.
*RST:
AUTO
Example:
DISP:MTAB ON
Activates the marker table.
Manual operation:
See
DISPlay[:WINDow<n>]:MINFo[:STATe]
<State>
This command turns the marker information in all diagrams on and off.
Suffix:
<n>
.
irrelevant
Parameters:
<State>
ON | 1
Displays the marker information in the diagrams.
OFF | 0
Hides the marker information in the diagrams.
*RST:
1
Example:
DISP:MINF OFF
Hides the marker information.
Manual operation:
See
CALCulate<n>:MARKer<m>:X:SSIZe
<StepSize>
This command selects the marker step size mode for
all
markers in
all
windows.
The step size defines the distance the marker moves when you move it with the rotary
knob.
It therefore takes effect in manual operation only.
Suffix:
<n>
.
irrelevant
Setting Basic Spectrum RF Measurement Parameters