Remote Commands
R&S
®
FPL1000
674
User Manual 1178.3370.02 ─ 09
[SENSe:]ESPectrum<sb>:RANGe<ri>:POINts:MINimum[:VALue]
<SweepPoint>
Defines the minimum number of sweep points for the range.
Suffix:
<sb>
.
1 to 3
<ri>
Selects the measurement range.
Parameters:
<SweepPoint>
Minimum number of sweep points per range
Range:
1 to 32001
*RST:
1
Example:
SENSe1:ESPectrum:RANGe3:POINts:MIN:VALue 400
Manual operation:
See
[SENSe:]ESPectrum<sb>:RANGe<ri>:MLCalc
<Function>
Defines the function used to calculate the limit line for the n-th power class for overlap-
ping ranges in Multi-SEM measurements. For details see
"Limit calculation for individ-
Suffix:
<sb>
.
1 to 3
Sub block in a Multi-SEM measurement
<ri>
1..n
Selects the measurement range.
Parameters:
<Function>
NONE | MAX | SUM
NONE
(reference ranges only:) the limit of the reference range is used;
Reference ranges always use the function "NONE".
SUM
sum of the two limit lines (calculated for linear powers) is used
MAX
maximum of the two limit lines is used
*RST:
SUM (reference range: NONE)
Manual operation:
See
[SENSe:]ESPectrum<sb>:RANGe<ri>:RLEVel
<RefLevel>
This command defines the reference level for a SEM range.
In case of high speed measurements, the reference level has to be identical for all
ranges.
Suffix:
<sb>
.
1 to 3
Sub block in a Multi-SEM measurement
Configuring and Performing Measurements