The Spectrum Application (RF Measurements)
R&S
®
FPL1000
396
User Manual 1178.3370.02 ─ 09
"Power"
Activates linear power averaging.
The power level values are converted into unit Watt prior to averag-
ing. After the averaging, the data is converted back into its original
unit.
Use this mode to average power values in Volts or Amperes correctly.
In particular, for small VBW values (smaller than the RBW), use
power averaging mode for correct power measurements in FFT
sweep mode.
Remote command:
Average Count
Determines the number of averaging or maximum search procedures If the trace
modes "Average", "Max Hold" or "Min Hold" are set.
In continuous sweep mode, if sweep count = 0 (default), averaging is performed over
10 sweeps. For sweep count =1, no averaging, Max Hold or Min Hold operations are
performed.
This value is identical to the
setting in the "Sweep" settings.
Remote command:
Predefined Trace Settings - Quick Config
Commonly required trace settings have been predefined and can be applied very
quickly by selecting the appropriate button.
Function
Trace Settings
Preset All Traces
Trace 1:
Clear Write
Auto Detector (Auto Peak)
Traces 2-6:
Blank
Auto Detector
Set Trace Mode
Max | Avg | Min
Trace 1:
Max Hold
Auto Detector (Positive Peak)
Trace 2:
Average
Auto Detector (Sample)
Trace 3:
Min Hold
Auto Detector (Negative Peak)
Traces 4-6:
Blank
Auto Detector
Set Trace Mode
Max | ClrWrite | Min
Trace 1:
Max Hold
Auto Detector (Positive Peak)
Trace 2:
Clear Write
Auto Detector (Auto Peak)
Trace Configuration