The Spectrum Application (RF Measurements)
R&S
®
FPL1000
173
User Manual 1178.3370.02 ─ 09
Figure 7-30: Summarized limit line for multiple sub blocks
7.2.7.5
SEM Configuration
Access
: "Overview" > "Select Measurement" > "Spectrum Emission Mask"
The SEM measurement is started immediately with the default settings.
The remote commands required to perform these tasks are described in
ter 11.6.7, "Measuring the Spectrum Emission Mask"
Global span settings
The span of the signal to be monitored is configured in the general span settings (see
Chapter 7.4.2, "Frequency and Span Settings"
on page 302). Only ranges within this
global span are considered for the SEM measurement.
Multi-SEM configuration
In the Spectrum application application only, spectrum emissions can be measured for
multiple sub blocks of channels (see
"SEM with Multiple Sub Blocks ("Multi-SEM")"
on page 169). Up to 3 sub blocks (with 2 gaps) can be defined. For each sub block, the
familiar configuration settings concerning ranges, limit lines etc. can be defined in indi-
vidual tabs. In addition, settings on the sub blocks themselves must be configured in
the "Sub Block" tab of the "Spectrum Emission Mask" configuration dialog box (see
"Multi-SEM (Sub Block) Settings "
The following settings are available in individual tabs of the "Spectrum Emission Mask"
configuration dialog box.
Measurements and Results