The Spectrum Application (RF Measurements)
R&S
®
FPL1000
363
User Manual 1178.3370.02 ─ 09
The remote commands required to define these settings are described in
ter 11.8.8, "Working with Markers"
Precise Frequency (Signal Count) Marker
............................................................363
Measuring Noise Density (Noise Meas Marker)
....................................................365
Phase Noise Measurement Marker
.......................................................................367
Measuring Characteristic Bandwidths (n dB Down Marker)
..................................370
........................................................................................372
Measuring the Power in a Channel (Band Power Marker)
....................................373
Demodulating Marker Values and Providing Audio Output (Marker Demodulation)
Deactivating All Marker Functions
.........................................................................382
7.9.4.1
Precise Frequency (Signal Count) Marker
Access
: "Overview" > "Analysis" > "Marker Functions" > "Select Marker Function" >
"Signal Count" > "Signal Count Config"
Or
: [MKR] > "Select Marker Function" > "Signal Count" > "Signal Count Config"
The frequency count marker is not available if the tracking generator is active (see
A normal marker determines the position of the point on the trace and indicates the sig-
nal frequency at this position. The trace, however, contains only a limited number of
points. Depending on the selected span, each trace point can contain many measure-
ment values. Thus, the frequency resolution of each trace point is limited.
(See also
Chapter 7.6.1.8, "How Much Data is Measured: Sweep Points and Sweep
Marker Usage