The Spectrum Application (RF Measurements)
R&S
®
FPL1000
108
User Manual 1178.3370.02 ─ 09
Figure 7-7: Measurement of two equally-leveled RF sinusoidal signals with a resolution band-
width which is larger than their frequency spacing
6. Set the resolution bandwidth to
1
kHz
.
The two generator signals are shown with high resolution. However, the sweep
time becomes longer. At smaller bandwidths, the noise display decreases simulta-
neously (10
dB decrease in noise floor for a decrease in bandwidth by a factor of
10).
Figure 7-8: Measurement of two equally-leveled RF sinusoidal signals with a resolution band-
width (1
kHz) which is significantly smaller than their frequency spacing
Measurements and Results