Remote Commands
R&S
®
FPL1000
782
User Manual 1178.3370.02 ─ 09
*OPC should be used after requesting data. This will hold off any subsequent changes
to the selected trigger source, until after the sweep is completed and the data is
returned.
Configuring the Triggering Conditions
...................................................................782
Configuring Gated Measurements
........................................................................ 786
11.8.4.1
Configuring the Triggering Conditions
The following commands are required to configure a triggered measurement.
...........................................................................................782
TRIGger[:SEQuence]:HOLDoff[:TIME]
..............................................................................782
TRIGger[:SEQuence]:IFPower:HOLDoff
........................................................................... 783
TRIGger[:SEQuence]:IFPower:HYSTeresis
.......................................................................783
TRIGger[:SEQuence]:LEVel[:EXTernal<port>]
...................................................................784
TRIGger[:SEQuence]:LEVel:IFPower
............................................................................... 784
TRIGger[:SEQuence]:LEVel:IQPower
...............................................................................784
.......................................................................................... 785
........................................................................................785
TRIGger[:SEQuence]:TIME:RINTerval
..............................................................................786
TRIGger[:SEQuence]:DTIMe
<DropoutTime>
Defines the time the input signal must stay below the trigger level before a trigger is
detected again.
Parameters:
<DropoutTime>
Dropout time of the trigger.
Range:
0 s to 10.0 s
*RST:
0 s
Default unit: S
Manual operation:
See
TRIGger[:SEQuence]:HOLDoff[:TIME]
<Offset>
Defines the time offset between the trigger event and the start of the sweep.
A negative offset is possible for time domain measurements.
For the trigger sources "External" or "IF Power", a common input signal is used for both
trigger and gate. Therefore, changes to the gate delay will affect the trigger offset as
well.