Remote Commands
R&S
®
FPL1000
779
User Manual 1178.3370.02 ─ 09
11.8.3.3
Configuring a Preamplifier
INPut<ip>:GAIN:STATe
<State>
This command turns the internal preamplifier on and off. It requires the optional pream-
plifier hardware.
If activated, the input signal is amplified by 20
dB. The preamplifier is only active below
3
GHz (R&S
FPL1003) or 7.5
GHz (R&S
FPL1007).
Suffix:
<ip>
.
1 | 2
irrelevant
Parameters:
<State>
ON | OFF | 0 | 1
OFF | 0
Switches the function off
ON | 1
Switches the function on
*RST:
0
Example:
INP:GAIN:STAT ON
Switches on 20 dB preamplification.
Manual operation:
See
11.8.3.4
Scaling the Y-Axis
DISPlay[:WINDow<n>]:TRACe<t>:Y[:SCALe]
................................................................... 779
DISPlay[:WINDow<n>]:TRACe<t>:Y[:SCALe]:AUTO ONCE
............................................... 780
DISPlay[:WINDow<n>][:SUBWindow<w>]:TRACe<t>:Y[:SCALe]:MODE
..............................780
DISPlay[:WINDow<n>]:TRACe<t>:Y[:SCALe]:RPOSition
................................................... 780
DISPlay[:WINDow<n>][:SUBWindow<w>]:TRACe<t>:Y:SPACing
........................................781
DISPlay[:WINDow<n>]:TRACe<t>:Y[:SCALe]
<Range>
This command defines the display range of the y-axis (for all traces).
Note that the command works only for a logarithmic scaling. You can select the scaling
with
DISPlay[:WINDow<n>][:SUBWindow<w>]:TRACe<t>:Y:SPACing
.
Suffix:
<n>
.
<t>
irrelevant
Parameters:
<Range>
Range:
1 dB to 200 dB
*RST:
100 dB
Default unit: HZ
Setting Basic Spectrum RF Measurement Parameters