Manual operation:
See
[SENSe:]SWEep:TIME
<Time>
This command defines the sweep time. It automatically decouples the time from any
other settings.
In the Spectrum application, the command decouples the sweep time from the span
and resolution and video bandwidths. Note that this command queries only the time
required to capture the data, not to process it. To obtain an estimation of the total cap-
ture and processing time, use the
command.
Parameters:
<Time>
refer to data sheet
*RST:
depends on current settings (determined automati-
cally)
Default unit: S
Manual operation:
See
See
See
[SENSe:]SWEep:TIME:AUTO
<State>
This command couples and decouples the sweep time to the span and the resolution
and video bandwidths.
Parameters:
<State>
ON | OFF | 0 | 1
*RST:
1
Example:
SWE:TIME:AUTO ON
Activates automatic sweep time.
Manual operation:
See
See
See
[SENSe:]SWEep:TYPE:USED
This command queries the sweep type if you have turned on automatic selection of the
sweep type.
Return values:
<Type>
SWE
Normal sweep
FFT
FFT mode