![Ronde & Schwarz 1304.0004.03 Скачать руководство пользователя страница 760](http://html1.mh-extra.com/html/ronde-and-schwarz/1304-0004-03/1304-0004-03_user-manual_1484985760.webp)
Remote Commands
R&S
®
FPL1000
760
User Manual 1178.3370.02 ─ 09
CALCulate<n>:MARKer<m>:FUNCtion:CSTep
This command matches the center frequency step size to the current marker fre-
quency.
The command turns delta markers into normal markers.
Suffix:
<n>
.
<m>
DISPlay[:WINDow<n>][:SUBWindow<w>]:TRACe<t>:X:SPACing
<Scale>
This command selects the scaling of the x-axis.
Suffix:
<n>
.
<w>
subwindow
<t>
Parameters:
<Scale>
LOGarithmic
Logarithmic scaling.
LINear
Linear scaling.
*RST:
LINear
Example:
DISP:TRAC:X:SPAC LOG
Manual operation:
See
[SENSe:]FREQuency:CENTer
<Frequency>
This command defines the center frequency.
Parameters:
<Frequency>
The allowed range and f
max
is specified in the data sheet.
UP
Increases the center frequency by the step defined using the
DOWN
Decreases the center frequency by the step defined using the
*RST:
fmax/2
Default unit: Hz
Example:
FREQ:CENT 100 MHz
FREQ:CENT:STEP 10 MHz
FREQ:CENT UP
Sets the center frequency to 110 MHz.
Setting Basic Spectrum RF Measurement Parameters