![Ronde & Schwarz 1304.0004.03 Скачать руководство пользователя страница 761](http://html1.mh-extra.com/html/ronde-and-schwarz/1304-0004-03/1304-0004-03_user-manual_1484985761.webp)
Remote Commands
R&S
®
FPL1000
761
User Manual 1178.3370.02 ─ 09
Manual operation:
See
See
[SENSe:]FREQuency:CENTer:STEP
<StepSize>
This command defines the center frequency step size.
You can increase or decrease the center frequency quickly in fixed steps using the
SENS:FREQ UP
AND
SENS:FREQ DOWN
Parameters:
<StepSize>
f
max
is specified in the data sheet.
Range:
1 to fMAX
*RST:
0.1 x span
Default unit: Hz
Example:
//Set the center frequency to 110 MHz.
FREQ:CENT 100 MHz
FREQ:CENT:STEP 10 MHz
FREQ:CENT UP
Manual operation:
See
[SENSe:]FREQuency:CENTer:STEP:AUTO
<State>
This command couples or decouples the center frequency step size to the span.
In time domain (zero span) measurements, the center frequency is coupled to the
RBW.
Parameters:
<State>
ON | OFF | 0 | 1
*RST:
1
Example:
FREQ:CENT:STEP:AUTO ON
Activates the coupling of the step size to the span.
[SENSe:]FREQuency:CENTer:STEP:LINK
<CouplingType>
This command couples and decouples the center frequency step size to the span or
the resolution bandwidth.
Parameters:
<CouplingType>
SPAN | RBW | OFF
SPAN
Couples the step size to the span. Available for measurements
in the frequency domain.
RBW
Couples the step size to the resolution bandwidth. Available for
measurements in the time domain.
Setting Basic Spectrum RF Measurement Parameters