![Ronde & Schwarz 1304.0004.03 Скачать руководство пользователя страница 904](http://html1.mh-extra.com/html/ronde-and-schwarz/1304-0004-03/1304-0004-03_user-manual_1484985904.webp)
Remote Commands
R&S
®
FPL1000
904
User Manual 1178.3370.02 ─ 09
CALCulate<n>:LIMit<li>:CONTrol:SHIFt
<Distance>
This command moves a complete limit line horizontally.
Compared to defining an offset, this command actually changes the limit line definition
points by the value you define.
Suffix:
<n>
.
irrelevant
<li>
Parameters:
<Distance>
Numeric value.
The unit depends on the scale of the x-axis.
Default unit: HZ
Manual operation:
See
CALCulate<n>:LIMit<li>:CONTrol:SPACing
<InterpolMode>
This command selects linear or logarithmic interpolation for the calculation of limit lines
from one horizontal point to the next.
Suffix:
<n>
.
<li>
Parameters:
<InterpolMode>
LINear | LOGarithmic
*RST:
LIN
Example:
CALC:LIM:CONT:SPAC LIN
Manual operation:
See
CALCulate<n>:LIMit<li>:LOWer[:DATA]
<LimitLinePoints>
This command defines the vertical definition points of a lower limit line.
Suffix:
<n>
.
irrelevant
<li>
Setting Basic Spectrum RF Measurement Parameters