![Ronde & Schwarz 1304.0004.03 Скачать руководство пользователя страница 212](http://html1.mh-extra.com/html/ronde-and-schwarz/1304-0004-03/1304-0004-03_user-manual_1484985212.webp)
The Spectrum Application (RF Measurements)
R&S
®
FPL1000
212
User Manual 1178.3370.02 ─ 09
File contents
Explanation
PeaksPerRange;25;
Max. number of peaks per range to be detected
Values;3;
Number of detected peaks
File data section
0;9000;150000;1000;79500;-25.006643295288086;-
12.006643295288086;PASS;
0;9000;150000;1000;101022.11126961483;-47.075
111389160156;-34.075111389160156;PASS;
0;9000;150000;1000;58380.171184022824;-47.079
341888427734;-34.079341888427734;PASS;
Measured peak values:
<range number>;
<start frequency>;
<stop frequency>;
<resolution bandwidth of range>;
<frequency of peak>;
<absolute power in dBm of peak>;
<distance to the limit line in dB>; (positive value
means above the limit)
<limit fail (pass = 0, fail =1)>;
7.2.9
Statistical Measurements (APD, CCDF)
To measure the amplitude distribution, the R&S
FPL1000 has simple measurement
functions to determine both the Amplitude Probability Distribution (APD) and the Com-
plementary Cumulative Distribution Function (CCDF). Only one of the signal statistic
functions can be switched on at a time.
......................................................................................212
...............................................................................................213
........................................................................................213
APD and CCDF Basics - Gated Triggering
........................................................... 215
.............................................................................. 216
How to Perform an APD or CCDF Measurement
..................................................222
Optimizing and Troubleshooting the Measurement
.............................................. 226
7.2.9.1
About the Measurements
The probability of amplitude values can be measured with the Amplitude Probability
Distribution function (APD). During a selectable measurement time all occurring ampli-
tude values are assigned to an amplitude range. The number of amplitude values in
the individual ranges is counted and the result is displayed as a histogram.
Alternatively, the Complementary Cumulative Distribution Function (CCDF) can be dis-
played. It shows the probability that the mean signal power amplitude will be exceeded
in percent.
Only one of the signal statistic functions can be switched on at a time. When a statistic
function is switched on, the R&S
FPL1000 is set into zero span mode automatically.
The R&S
FPL1000 measures the statistics of the signal applied to the RF input with
the defined analysis bandwidth. To avoid affecting the peak amplitudes the video band-
Measurements and Results