Remote Commands
R&S
®
FPL1000
676
User Manual 1178.3370.02 ─ 09
[SENSe:]ESPectrum<sb>:RANGe<ri>:TRANsducer
<Transducer>
This command selects a transducer factor for a SEM range.
Note that
●
the transducer must cover at least the span of the range
●
the x-axis has to be linear
●
the unit has to be dB
Suffix:
<sb>
.
1 to 3
Sub block in a Multi-SEM measurement
<ri>
1..n
Selects the measurement range.
Parameters:
<Transducer>
String containing the transducer file name, including the path
information.
Manual operation:
See
[SENSe:]ESPectrum<sb>:SSETup
<State>
Enables or disables symmetrical configuration of the range settings.
See
Suffix:
<sb>
.
1 to 3
Parameters:
<State>
ON | OFF | 0 | 1
OFF | 0
Switches the function off
ON | 1
Switches the function on
*RST:
0
Manual operation:
See
11.6.7.5
Configuring the Reference Range
The following commands define the reference range for the SEM sweep list.
...................................................................................... 677
[SENSe:]ESPectrum<sb>:FILTer[:RRC]:ALPHa
.................................................................677
[SENSe:]ESPectrum<sb>:FILTer[:RRC][:STATe]
................................................................677
................................................................................677
.................................................................................... 678
Configuring and Performing Measurements