Data Management
R&S
®
FPL1000
458
User Manual 1178.3370.02 ─ 09
Saving and recalling transducer and limit line settings
If a transducer file was in use when the save set was stored (with the save item "Cur-
rent Settings" only) the R&S
FPL1000 assumes that these transducer values should
remain valid after every recall of that save set. Thus, even if the transducer file is
changed and the original save set file is recalled later, the
originally stored
transducer
values are recalled and applied to the measurement. In the "Edit" transducer dialog
box, however, the
changed
transducer file values are displayed, as no updated trans-
ducer file was loaded.
The same applies to limit line settings.
The same applies to integrated measurements' weighting filter.
If you want to apply the changed transducer values after recalling the save set, you
must force the application to reload the transducer file. To do so, simply open the "Edit
Transducer" dialog box (see
Chapter 9.7.2, "Transducer Settings"
on page 525) and
toggle the "X-Axis" option from "Lin" to "log" and back. Due to that change, the trans-
ducer file is automatically reloaded, and the changed transducer values are applied to
the current measurement. Now you can create a new save set with the updated trans-
ducer values.
Similarly, if you want to apply the changed limit values after recalling the save set, you
must force the application to reload the limit file. To do so, simply open the "Edit Limit
Line" dialog box (see
Chapter 7.11.2.2, "Limit Line Settings and Functions"
on page 425) and toggle the "Y-Axis" unit. Due to that change, the limit line file is auto-
matically reloaded, and the changed limit values are applied to the current measure-
ment. Now a new save set with the updated limit values can be created.
.......................................................................................458
Configurable Storage and Recall
.......................................................................... 460
How to Save and Load Instrument Settings
..........................................................465
8.2.1
Quick Save/Quick Recall
The "Quick Save" and "Quick Recall" functions allow you to store instrument settings or
channel setups very easily and quickly in one step. Up to ten different sets of settings
can be stored to or recalled from "save sets". Each save set is identified by its storage
date and type (instrument or specific "Channel") in the display. The save sets are
stored in the
C:\Users\Public\Documents\Rohde-Schwarz\Analyzer\QuickSave
direc-
tory, in files named
QuickSave1.dfl
to
QuickSave10.dfl
. Only the current mea-
surement settings are stored, not any additional data such as traces, limit line or trans-
ducer files (see
Chapter 8.2.2.1, "Stored Data Types"
During recall, save sets of type "Instrument" replace the settings of the entire instru-
ment. All other save sets start a new channel setup with the stored settings.
Storing and Recalling Instrument Settings and Measurement Data