![Ronde & Schwarz 1304.0004.03 Скачать руководство пользователя страница 808](http://html1.mh-extra.com/html/ronde-and-schwarz/1304-0004-03/1304-0004-03_user-manual_1484985808.webp)
Remote Commands
R&S
®
FPL1000
808
User Manual 1178.3370.02 ─ 09
11.8.5.4
Configuring the Outputs
The following commands are required to provide output from the R&S
FPL1000.
Output functions require the option R&S
FPL1-B5 to be installed on the R&S
FPL1000.
........................................................................................808
..............................................................................................808
......................................................................................... 809
............................................................................................809
..........................................................................................809
OUTPut:UPORt:WTRigger:POLarity
................................................................................ 810
.............................................................................................. 810
................................................................................................ 811
.............................................................................................811
DIAGnostic:SERVice:NSOurce
<State>
This command turns the 28 V supply of the BNC connector labeled [noise source
control] on the R&S
FPL1000 on and off.
Parameters:
<State>
ON | OFF | 0 | 1
OFF | 0
Switches the function off
ON | 1
Switches the function on
Example:
DIAG:SERV:NSO ON
Manual operation:
See
OUTPut<up>:IF[:SOURce]
<Source>
Defines the type of signal available at one of the output connectors of the
R&S
FPL1000.
Suffix:
<up>
.
Parameters:
<Source>
IF
The measured IF value is available at the IF/VIDEO output con-
nector.
This connector is only available if the R&S
FPL1-B5 option is
installed.
The frequency at which the IF value is provided is defined using
the
VIDeo
The displayed video signal (i.e. the filtered and detected IF sig-
nal, 200mV) is available at the IF/VIDEO output connector.
This setting is required to provide demodulated audio frequen-
cies at the output.
Setting Basic Spectrum RF Measurement Parameters