Remote Commands
R&S
®
FPL1000
744
User Manual 1178.3370.02 ─ 09
Parameters:
<TimeOffset>
Defines a time offset to start the measurement at the first pulse
of a trace.
*RST:
0
Default unit: s
<MeasTime>
Defines the measurement time.
Default unit: s
<PulsePeriod>
Defines the pulse period.
Default unit: s
<OfPulses>
Defines the number of pulses to measure.
Example:
CALC:MARK:FUNC:MSUM 50US,450US,576.9US,8
Evaluates data that contains 8 pulses during a measurement
time of 450
µs and a pulse period of 576.9
µs. The evaluation
starts with an offset of 50
µs.
[SENSe:]MPOWer:FTYPe
<FilterType>
This command selects the filter type for pulse power measurements.
Parameters:
<FilterType>
CFILter
NORMal
[SENSe:]MPOWer:RESult[:LIST]?
This command queries the results of the pulse power measurement.
This command may be used to obtain measurement results in an asynchronous way,
using the service request mechanism for synchronization to the end of the measure-
ment.
If there are no results, the command returns an error.
Return values:
<PulsePower>
List of pulse powers.
The number of values depends on the number of pulses you
have been measuring.
The unit is dBm.
Usage:
Query only
[SENSe:]MPOWer[:SEQuence]
<Frequency>, <RBW>, <MeasTime>,
<TriggerSource>, <TriggerLevel>, <TriggerOffset>, <Detector>, <NoPulses>
This command configures and initiates the pulse power measurement.
Configuring and Performing Measurements