![Ronde & Schwarz 1304.0004.03 Скачать руководство пользователя страница 665](http://html1.mh-extra.com/html/ronde-and-schwarz/1304-0004-03/1304-0004-03_user-manual_1484985665.webp)
Remote Commands
R&S
®
FPL1000
665
User Manual 1178.3370.02 ─ 09
Usage:
Query only
[SENSe:]ESPectrum<sb>:RANGe<ri>:DELete
This command removes a range from the sweep list.
Note that
●
you cannot delete the reference range
●
a minimum of three ranges is mandatory.
Suffix:
<sb>
.
1 to 3
<ri>
Selects the measurement range.
Manual operation:
See
[SENSe:]ESPectrum<sb>:RANGe<ri>:FILTer:TYPE
<FilterType>
This command selects the filter type for an SEM range.
In case of high speed measurements, the filter has to be identical for all ranges.
The EMI-specific filter types are available if the EMI (R&S
FPL1-K54) measurement
option is installed, even if EMI measurement is not active. For details see
Suffix:
<sb>
.
1 to 3
<ri>
1...30
Selects the measurement range.
Parameters:
<FilterType>
NORMal
Gaussian filters
CFILter
channel filters
RRC
RRC filters
CISPr | PULSe
CISPR (6 dB) - requires EMI (R&S
FPL1-K54) option
Return value for query is always
PULS
.
MIL
MIL Std (6 dB) - requires EMI (R&S
FPL1-K54) option
P5
5 Pole filters
*RST:
NORM
Refer to the datasheet for available filter bandwidths.
Manual operation:
See
Configuring and Performing Measurements