Remote Commands
R&S
®
FPL1000
818
User Manual 1178.3370.02 ─ 09
<w>
subwindow
<t>
Parameters:
<Aperture>
Range:
1 to 50
*RST:
2
Default unit: PCT
Example:
DISP3:TRAC2:SMO:APER 5
Defines an aperture of 5% for trace 2 in window 3
Manual operation:
See
DISPlay[:WINDow<n>][:SUBWindow<w>]:TRACe<t>:SMOothing[:STATe]
<State>
This command turns trace smoothing for a particular trace on and off.
If enabled, the trace is smoothed by the value specified using
WINDow<n>][:SUBWindow<w>]:TRACe<t>:SMOothing:APERture
For more information see
Suffix:
<n>
.
<w>
subwindow
<t>
Parameters:
<State>
ON | OFF | 0 | 1
OFF | 0
Switches the function off
ON | 1
Switches the function on
Example:
DISP3:TRAC2:SMO ON
Turns on trace smoothing for trace 2 in window 3
Manual operation:
See
[SENSe:]AVERage<n>:COUNt
<AverageCount>
This command defines the number of sweeps that the application uses to average
traces.
In case of continuous sweep mode, the application calculates the moving average over
the average count.
In case of single sweep mode, the application stops the measurement and calculates
the average after the average count has been reached.
Suffix:
<n>
.
irrelevant