Remote Commands
R&S
®
FPL1000
706
User Manual 1178.3370.02 ─ 09
●
Reference level (x-axis)
0.0 dBm
●
Display range (x-axis) for APD measurements
100 dB
●
Display range (x-axis) for CCDF measurements
20 dB
●
Upper limit of the y-axis
1.0
●
Lower limit of the y-axis
1E-6
Suffix:
<n>
.
Example:
CALC:STAT:PRES
Resets the scaling for statistical functions
Manual operation:
See
CALCulate<n>:STATistics:SCALe:AUTO ONCE
This command initiates an automatic scaling of the diagram (x- and y-axis).
To obtain maximum resolution, the level range is set as a function of the measured
spacing between peak power and the minimum power for the APD measurement and
of the spacing between peak power and mean power for the CCDF measurement. In
addition, the probability scale for the number of test points is adapted.
To get valid results, you have to perform a complete sweep with synchronization to the
end of the auto range process. This is only possible in single sweep mode.
Suffix:
<n>
.
Manual operation:
See
CALCulate<n>:STATistics:SCALe:X:RANGe
<Range>
This command defines the display range of the x-axis for statistical measurements.
The effects are identical to
DISPlay[:WINDow<n>]:TRACe<t>:Y[:SCALe]
Suffix:
<n>
.
Parameters:
<Range>
Range:
1 dB to 200 dB
*RST:
100 dB
Default unit: dB
Example:
CALC:STAT:SCAL:X:RANG 20dB
Manual operation:
See
Configuring and Performing Measurements