![Ronde & Schwarz 1304.0004.03 Скачать руководство пользователя страница 350](http://html1.mh-extra.com/html/ronde-and-schwarz/1304-0004-03/1304-0004-03_user-manual_1484985350.webp)
The Spectrum Application (RF Measurements)
R&S
®
FPL1000
350
User Manual 1178.3370.02 ─ 09
Assigning the Marker to a Trace
................................................................................. 351
Selected Marker
Marker name. The marker which is currently selected for editing is highlighted orange.
Remote command:
Marker selected via suffix <m> in remote commands.
Marker State
Activates or deactivates the marker in the diagram.
Remote command:
CALCulate<n>:MARKer<m>[:STATe]
CALCulate<n>:DELTamarker<m>[:STATe]
Marker Position X-value
Defines the position (x-value) of the marker in the diagram. For normal markers, the
absolute position is indicated. For delta markers, the position relative to the reference
marker is provided.
Remote command:
Frame (Spectrogram only)
Spectrogram frame the marker is assigned to.
Remote command:
CALCulate<n>:MARKer<m>:SPECtrogram:FRAMe
CALCulate<n>:DELTamarker<m>:SPECtrogram:FRAMe
Marker Type
Toggles the marker type.
The type for marker 1 is always "Normal", the type for delta marker 1 is always "Delta".
These types cannot be changed.
Note:
If normal marker 1 is the active marker, switching the "Mkr Type" activates an
additional delta marker 1. For any other marker, switching the marker type does not
activate an additional marker, it only switches the type of the selected marker.
"Normal"
A normal marker indicates the absolute value at the defined position
in the diagram.
"Delta"
A delta marker defines the value of the marker relative to the speci-
fied reference marker (marker 1 by default).
Remote command:
CALCulate<n>:MARKer<m>[:STATe]
CALCulate<n>:DELTamarker<m>[:STATe]
Reference Marker
Defines a marker as the reference marker which is used to determine relative analysis
results (delta marker values).
Marker Usage