The Spectrum Application (RF Measurements)
R&S
®
FPL1000
178
User Manual 1178.3370.02 ─ 09
If the function is set to "Max", you can define a relative
and
an absolute limit level. In
this case, the maximum of the two values is used as the limit level.
For more information, see
"Relative limit line functions"
Remote command:
[SENSe:]ESPectrum<sb>:RANGe<ri>:LIMit<li>:RELative:STARt
[SENSe:]ESPectrum<sb>:RANGe<ri>:LIMit<li>:RELative:STOP
[SENSe:]ESPectrum<sb>:RANGe<ri>:LIMit<li>:RELative:STARt:
FUNCtion
[SENSe:]ESPectrum<sb>:RANGe<ri>:LIMit<li>:RELative:STOP:FUNCtion
[SENSe:]ESPectrum<sb>:RANGe<ri>:LIMit<li>:RELative:STARt:ABS
[SENSe:]ESPectrum<sb>:RANGe<ri>:LIMit<li>:RELative:STOP:ABS
Multi-Limit Calc <n>
Defines the function used to calculate the limit line for the n-th power class for overlap-
ping ranges in Multi-SEM measurements. For details, see
"Limit calculation for individ-
"NONE"
(reference ranges only:) the limit of the reference range is used
"SUM"
Sum of the two limit lines (calculated for linear powers) is used
"MAX"
Maximum of the two limit lines is used
Remote command:
[SENSe:]ESPectrum<sb>:RANGe<ri>:MLCalc
Min Sweep Points
Defines the minimum number of sweep points for the range.
If necessary to fulfill all minimum sweep point requirements in all ranges, the global
setting is increased. By default, each range is supplied with a minimum
of one sweep point.
For details, see
Remote command:
[SENSe:]ESPectrum<sb>:RANGe<ri>:POINts:MINimum[:VALue]