Remote Commands
R&S
®
FPL1000
805
User Manual 1178.3370.02 ─ 09
DISPlay[:WINDow<n>][:SUBWindow<w>]:TRACe<t>:Y[:SCALe]:RVALue
<Value>
This command defines the reference value assigned to the reference position in the
specified window. Separate reference values are maintained for the various displays.
Suffix:
<n>
.
<w>
subwindow
<t>
irrelevant
Parameters:
<Value>
*RST:
AM time domain: 0 PCT; FM time domain: 0 Hz; PM
time domain: 0 rad; AM spectrum: 100 PCT; FM
spectrum: 250 kHz; PM spectrum: 10 rad;
Default unit: DB
Example:
DISP:TRAC:Y:RVAL 0
Sets the value assigned to the reference position to 0 Hz
Manual operation:
See
Power Sweep Commands
The following commands are required to configure a power sweep using an internal
tracking generator.
For an example of performing a power sweep remotely, see
gramming Example: Measuring Compression Points"
Useful commands for power sweeps described elsewhere:
●
●
DISPlay[:WINDow<n>]:TRACe<t>:Y[:SCALe]:RPOSition
●
SOURce<si>:INTernal:FREQuency:OFFSet
●
SOURce<si>:POWer[:LEVel][:IMMediate]:OFFSet
●
●
Remote commands exclusive to power sweeps:
........................................................................................... 805
............................................................................................806
............................................................................................ 806
SOURce<si>:POWer:MODE
<Mode>
Enables or disables a power sweep.
Suffix:
<si>
.
irrelevant
Parameters:
<Mode>
FIXed | SWEep
Setting Basic Spectrum RF Measurement Parameters