During testing, MADSEN OTOflex 100 and the Immittance module will show a probe icon indicating leakage, if any.
See
Leakage and other probe problems
7.4.4
Leakage and other probe problems
Testing may be complicated by a number of factors which can result in leakage or probe problems.
Leakage - possible causes:
•
The eartip does not fit well
•
The eartip is not inserted properly in the ear canal
•
The probe tip opening is blocked by the wall of the ear canal
•
The eartip may be old or hardened
•
the threaded ring holding the probe tip has not been tightened properly
•
The pneumatic probe plug has not been inserted properly in OTOflex
•
the probe tip has not been tightened properly
•
pneumatic probe plug not inserted properly in MADSEN OTOflex 100
Probe problems can be caused by:
•
an occluded probe
•
a blocked wax filter
Probe status icons
MADSEN OTOflex 100
OTOsuite
Description
The probe is OK
The probe is blocked
There is a probe leak
The probe is not fully inserted
7.5
Preparing for the next patient
When you have finished testing a patient and wish to test a new patient:
1. For the sake of patient privacy, when you receive a new patient make sure that the screen does not show any inform-
ation about the patient you just finished testing.
2. If you have set
Menu
>
Procedure options..
>
'Patient' prompt
to
On
, MADSEN OTOflex 100 will present the
Patient
and User
data screen as part of the
‘Done’
action sequence. See
3. Enter the patient’s name and date of birth to identify the data on the single page report. See
and test type in MADSEN OTOflex 100
56