1. If you want to manually change the automatically
determined threshold:
click on the drop-down box of the specific field in
the Thresholds table and select the stimulus intens-
ity to be reported as the threshold value. The manu-
ally selected value will be marked by an asterisk, the
appropriate graph at the top of the screen is framed
accordingly, and the audiogram marker in the audi-
ogram in the bottom right part of the screen is
updated.
2. If you want to continue with manual testing, see
11.6.3
Settings - Reflex Threshold
Setting
Value
Toolbar icon
Auto resume on seal
(
Tools
>
Options
>
General
> )
•
Off
:
If there is a probe leak, the test is interrupted. You must resume
testing manually if testing is paused due to a leak.
On
:
Testing resumes automatically when a proper seal is obtained if test-
ing was paused due to a leak.
Custom TPP offset
(
Control Panel
)
•
Checked
:
The pressure is offset from the TPP by the TPP offset defined in
Tools
>
Options
>
Reflex Threshold
>
General
>
Custom TPP
offset
. If the offset is is set to a value with the post fix “more”, it
shifts the pressure in the same direction as the sign of the TPP, and
in the opposite direction if it is set to “less”.
Note
•
The TPP value from the most recent tympanometric curve
at the same probe tone is used in reflex measurements.
Note
•
When the
Custom TPP offset
setting is changed in Acous-
tic Reflex Threshold, this setting is automatically applied in the
Acoustic Reflex Decay test setting to ensure consistent reflex
testing, but not vice-versa.