Audiogram
The audiogram marker in the audiogram in the bottom right part of the screen indicates the threshold level or the
highest intensity where threshold was not found.
5.4.3
The Reflex Decay screen
Field results
Deflection
The deflection graph is limited by blue markers
indicating stimulus on and off.
A third marker appears as indication of the half-life
time where the curve decreases to 50% of its initial
deflection.
Half-life time, seconds - Results table
•
If the deflection decreases to 50% or more of the initial deflection during presentation of the stimulus, the point in
time when this occurs is shown.
If no decay is registered, the field shows a dash.
The Ipsi results are shown in the top row. The stimulus ear referenced contra-lateral results appear in the bottom
row.
Tympanogram
In the top right corner of the screen, the most recent tympanogram, if available for the currently used probe tone, is
displayed, and a marker on the pressure axis indicates the current ear canal pressure.
The curve number is also indicated.