Setting
Value
Toolbar icon
Custom TPP offset
(
Control Panel
)
•
Checked
:
The pressure is offset from the TPP by the TPP offset defined in
Tools
>
Options
>
Reflex Threshold
>
General
>
Custom TPP
offset
. If the offset is is set to a value with the post fix “more”, it
shifts the pressure in the same direction as the sign of the TPP, and
in the opposite direction if it is set to “less”.
Note
•
The TPP value from the most recent tympanometric curve
at the same probe tone is used in reflex measurements.
Note
•
When the
Custom TPP offset
setting is changed in Acous-
tic Reflex Threshold, this setting is automatically applied in the
Acoustic Reflex Decay test setting to ensure consistent reflex
testing, but not vice-versa.
Verification
(
Tools
>
Options
>
Reflex Threshold
> )
•
Depending on your settings, you can select the type of verification to
be used after the deflection criterion has been met by the auto-
matic threshold search: The lowest stimulus intensity, where the
reflex criterion was met, may be retested for reproduceability, or
the next one or two intensity levels can be used to confirm the
reflex growth properties.
8.10
ETF-P (Eustachian Tube Function - Perforated)
8.10.1
Softkeys and functions
Idle mode functions
Softkeys
Functions
Left
Press and hold to display patient and user info
Right
Press to run test (clear curve and wait for pump directions)
Otometrics - MADSEN OTOflex 100
83
8 Testing with MADSEN OTOflex 100