Normal
•
The pressure range settings are disabled.
Custom range
is not
enabled.
•
The pressure range selected in the currently used measurement
setup is shown.
The normal range of pump pressure is from -400 daPa to
+200 daPa.
Custom range
•
Enable to access and change the values in the
Pressure range
fields.
•
You can extend the pressure range of –400 to +200 daPa to as
much as –600 to +400 daPa, or decrease it to -50 to +50 daPa in
steps of 50 daPa.
Default min. pressure: -600 daPa
Default max. pressure: 400 daPa
Sweep direction
The air pressure determines the direction of the pressure sweep
from a positive value to a negative value or vice versa.
•
Positive
- from a negative towards a positive value
•
Negative
- from a positive towards a negative value
Default:
Negative
14.3.1.2
Tympanometry: Test settings
Click icon or select
Tools
>
Options
>
Tymp. and Reflex Scr.
>
Measurement
Tympanometry
Auto next curve...
Enable or disable to define whether the system will automatically
select the next curve number available for the new measurement res-
ults when you press
Start
to start a new measurement.
The tympanometry graph shows a maximum of 3 curves per ear per
patient.
Otometrics - MADSEN OTOflex 100
151
14 Immittance module settings