14.1
General settings
These settings apply regardless of the test type or sequence selected.
1. To change general settings, select
Tools
>
Options
>
Immittance
>
General
or click the
Set Options
icon on the toolbar.
2. The
Options
dialog appears.
3. Click on
Immittance
>
General
.
Measurement settings
Auto Settings..
Auto resume on seal
If set to
On
, and a measurement is paused by a leak, the test or
sequence is automatically resumed when the probe is properly refit-
ted in the ear canal.
Auto start on seal
The system automatically starts the measurement as good seal is
obtained in the ear canal. Regardless of the setting in
Tools
>
Options
>
Tymp. and Reflex Scr.
>
Auto next curve...
, the Immit-
tance module automatically changes to the next tymp curve for each
tymp sweep. This is to prevent overwriting already recorded tymps, if
a tymp measurement is triggered by accident.
This setting is part of the measurement setting. Depending on the
selected test setting, define this setting to suit the specific test set-
ting purposes.
First ear to test
First ear to test
Defines the first ear to be tested on a new patient.
General
Settings Name
Enter the term/name to be used for settings saved as a setup in
MADSEN OTOflex 100.
Safety
Intensity Safety
On
:
When Intensity safety is on, no stimulus intensities higher than 115
dB SPL (measured in a 2cc cavity) can be selected or will be applied.
Off
:
Stimulus intensities up to the technical limits of the device and
probe can be selected and applied. A warning triangle is displayed
on-screen whenever intensities above 115 dB SPL are selected.
Default:
On