Restoring a faulty connection
1. Click the
Select Device
icon, or select
Tools
>
Select and Manage Test Devices..
2. Click the
Repair
button to repair a faulty connection to the highlighted test device. Click
Search
if the
repair function only made the test device disappear from the list.
For more information, see
Configuring the Immittance module
.
6.3
Deleting a test device
1. Click the
Select Device
icon, or select
Tools > Select and Manage Test Devices..
2. Right-click the device name and select
Delete
to remove the test device from the list. Removed
devices can be added again using the
Search
button.
6.4
Test device information
To see information relating to the test device, select
Help > About Device
.
6.5
Firmware upgrade - test devices
If your OTOsuite version contains a more recent firmware for your MADSEN OTOflex 100, it is recommended that you
update the firmware to make sure that MADSEN OTOflex 100 and OTOsuite perform correctly.
Upgrading of the firmware will leave your device settings unchanged.
The new firmware may take up slightly more MADSEN OTOflex 100 memory than the previous version. If the memory is
full of patient results, the upgrade process will automatically erase some of the oldest patient results from the MADSEN
OTOflex 100 memory if needed (results that have been saved or printed will be deleted first).
It is recommended that you manually delete some or all patient results from MADSEN OTOflex 100 to create the neces-
sary space for the upgrade.
Firmware upgrade procedure
1. Place MADSEN OTOflex 100 in a powered charger near the PC and
Bluetooth adapter.
2. Connect to the specific MADSEN OTOflex 100 you want to upgrade
from OTOsuite (as if you wanted to do testing).
3. Disconnect MADSEN OTOflex 100.
4. In OTOsuite select
Tools
>
Configuration Wizard
>
Configure
>
Immittance
>
Next
to go to the
Firmware Update
screen.
5. Click the
Upgrade OTOflex firmware
button.
Otometrics - MADSEN OTOflex 100
45
6 Communication between OTOsuite and MADSEN OTOflex 100