The test progresses as follows
•
If a threshold is not found immediately:
The stimulus intensities will increase until a reflex threshold is registered or until the max.
intensity is reached.
•
If a threshold is found immediately:
The test will automatically decrease the stimulus intensities until a threshold is no longer
registered.
•
Deflection curves
The deflection curves result in a complete reflex deflection graph. The curves shown are sor-
ted according to increased stimulus intensity.
•
Thresholds table
Shows the determined threshold. If no threshold is detected, the field in the Thresholds table
will show the text
None
to indicate the absent threshold.
•
Tympanogram
If available, the most recent tympanogram for the currently used probe tone is displayed, and a
marker on the pressure axis indicates the current ear canal pressure.
The curve number is also indicated.
The following symbols are used in the audiogram:
–
Pure tone air unmasked/masked
–
Pure tone bone unmasked/masked
–
Ipsi Thr. found/not found
–
Contra Thr. found/not found (Stim: x)
8.8.3
Viewing Reflex Threshold results
This Reflex Threshold test result example shows the three deflection curves for
the highlighted 1 kHz ipsilateral stimulus.
Manually changed thresholds are marked with an asterisk.
The dashed horizontal line in each reflex graph indicates the predefined reflex cri-
terion.
Repeating a measurement
If you repeat a measurement at a specific intensity, the previous measurement
for that intensity will be overwritten. You cannot see several curves with the
same intensity.
Reflex Threshold field results
•
If afield shows a value with no extra marks, the threshold has been determined automatically.
•
If afield is crossed out, this indicates that no threshold has been detected at this highest test level.
•
If afield is marked by an asterisk, the threshold has been manually determined.
•
Blank fields indicate that the test has not been performed.
76
Otometrics - MADSEN OTOflex 100
8 Testing with MADSEN OTOflex 100