14.3.6
ETF-P setup
14.3.6.1
ETF-P: Control Panel settings
ETF-P
Initial pressure
The pressure used for the first part of the test.
Positive pressure
Select the maximum positive pressure to be presented to the ear
canal/middle-ear cavity.
0 to 400 daPa, presented in 50 daPa increments.
Default: 400 daPa
Negative pressure
Select the maximum negative pressure to be presented to the ear
canal/middle-ear cavity.
-100 to -600 daPa, presented in 50 daPa increments.
Default: -400 daPa
14.3.6.2
ETF-P: Test settings
Click icon or select
Tools
>
Options
>
ETF-P
>
Measurement
.
Pump speed(daPa/s)
Pump speed(daPa/s)
Changes the speed of the pressure sweep. Indicated in daPa per
second.
Note
•
It is highly recommended that the lowest setting (50
daPa/s) is always used for ETF-P.
While the pump is operating, a scale at the top of the screen indic-
ates the current pump position graphically, i.e. the proportion of the
pump’s capacity being used to obtain the static pressure used for the
measurement.