Stimulus1
Menu
>
More settings..
500 Hz, 1000 Hz, 2000 Hz, 4000 Hz, BBN.
Default: 1000 Hz
Stimulus2
Menu
>
More settings..
500 Hz, 1000 Hz, 2000 Hz, 4000 Hz, BBN,
Off
.
Default: 2000 Hz
Stimulus3
Menu
>
More settings..
500 Hz, 1000 Hz, 2000 Hz, 4000 Hz, BBN,
Off
.
Default: 4000 Hz
Stimulus4
Menu
>
More settings..
500 Hz, 1000 Hz, 2000 Hz, 4000 Hz, BBN,
Off
.
Default: 500 Hz
Stimulus5
Menu
>
More settings..
500 Hz, 1000 Hz, 2000 Hz, 4000 Hz, BBN,
Off
.
Default:
Off
Stimulus side
Menu
>
More settings..
Ipsi, Contra, Ipsi then Contra, Contra then Ipsi (diagnostic version only).
Default:
Ipsi
Max intensity
Menu
>
More settings..
Select the desired maximum level of the stimulus intensity (up to 100 dB).
Default: 95 dB HL
The test is interrupted at Max level and must be continued manually.
Warning
•
Be careful when you enter the Max setting, so as not to expose the
patient to too high intensity levels!
Note
•
To avoid automatic testing being interrupted because of high stimulus
intensity levels when reaching the warning limits, it is recommended that you
set the max. intensity to 95 dB HL.
136
Otometrics - MADSEN OTOflex 100
13 Test settings in MADSEN OTOflex 100