Procedure
3. Press
On
/
Off
briefly to select the ETF-P test screen.
This screen shows results, if available, from a previous test.
4. Press the right-hand softkey to start the test.
5. Depending on the setting in
Menu
>
Initial pressure
, positive pressure auto-
matically builds up until the first opening of the Eustachian tube or max. pres-
sure is reached, as indicated by right softkey.
The pressure control is automatically reversed after one of the two criteria
has been reached.
If the initial time scale value is exceeded during the measurement, the axis
automatically expands by 10 second steps to a maximum of 60 seconds.
/
8.10.3
Viewing ETF-P results
The ETF-P test screen in MADSEN OTOflex 100 provides a simple view for quickly assessing the Eustachian tube function.
For more extensive analysis, use the ETF-P screen post-analysis tool OTOsuite Immittance module to retrieve the numer-
ical pressure values.
This ETF-P test result example shows how the pressure changes throughout the
measurement as the Eustachian tube briefly opens and closes.
One curve per ear is saved.
Changing view settings
You can change the following view setting to customize your view. This setting does not influence the test data as such.
•
Menu
>
More settings..
>
Initial time scale
Otometrics - MADSEN OTOflex 100
85
8 Testing with MADSEN OTOflex 100