Softkeys
Functions
Right
Start
stimulus. If safety intensity level is > safety limit: request confirmation to con-
tinue anyway
Stop
stimulus.
Scroll wheel
Select intensity. Flashing value indicates intensity warning (>108 dB SPL in 2 cc)
Select
Return to idle mode
8.8.2
Reflex Threshold - procedure
Reflex Threshold automatically follows the tympanometry sweep if you start the T+RT sequence (see
).
Note
•
To avoid automatic testing being interrupted because of high stimulus intensity levels when reaching the
warning limits, it is recommended that you set the max. intensity to 95 dB HL.
Procedure
1. Press
On
/
Off
briefly to select the Reflex Threshold test screen.
The initial reflex threshold test screen is shown with ipsilateral 1 kHz stim-
ulus highlighted.
The dashed horizontal line in each reflex graph indicates the predefined
reflex criterion.It is offset from the base line marker on the vertical axis. The
degree of offset is influenced by the setting in
Menu
>
More settings..
>
Reflex criterion
.
Warning
•
If the patient is troubled by the test, stop the test. The test is
interrupted and the pump pressure is relieved immediately.Already meas-
ured results are kept.
Automatic RT testing
–
Press the right-hand softkey to start a fully automatic reflex threshold
search for all preset stimuli.
The test runs a fully automatic threshold search for the preset stimuli
(
Menu
>
More settings..
>
Auto Settings..
).
74
Otometrics - MADSEN OTOflex 100
8 Testing with MADSEN OTOflex 100