Scale (Not com-
pensated)
The two different scale values separated by a slash (e.g. 3.0/6.0) are
probe tone dependent. The first value applies when 226 Hz probe
tone is used, the second applies when 1000 Hz is used.
When tympanograms are displayed with a scale of 1.5 or 3.0 mmho
(or cc/cm
3
/ml), the graph will show an axis complying with the stand-
ard ANSI aspect ratio, where 1 mmho corresponds to 300 daPa in
magnitude.
Auto Scale
icon
Enables automatic rescaling to a desired value in order to display the
entire curve.
Show norm area
For selecting/deselecting display of the norm area in the graph.
The modified Jerger normative data applies to 226 Hz only.
Select to enable the
Selected Norm Region
list.
Selected Norm
Region
Possible selection:
Adult, 226 Hz
, which is based on the Jerger norm.
For controlling whether auto classification is performed and according
to which norm.
The
Show norm Area
result is directly linked to the
Adult, 226 Hz
norm. This is because the tympanometric type classification is
defined by the TPP relation to the range of the norm data.
Admittance axis
Select
Baseline compensation
or
Not compensated
mode. This will
enable you to set the scaling of the view setting for
Scale (Baseline
compensated)
and
Scale (Not compensated)
individually.
•
Baseline compensated
Enabling
Baseline compensated
displays a tympanogram view
showing the compensated admittance curve as measured at the
tympanic membrane level.
The tympanometric peak in the graph will represent the estim-
ated admittance of the middle ear only. See the description of
Static Admittance
in the MADSEN OTOflex 100 documentation.
Select
Baseline compensated
to access the fields
Scale
(Baseline compensated)
,
B/G numeric component com-
pensation for Y
and
Show normal area
.
•
Not compensated
Select to display a tympanogram showing the total admittance
curve as measured at the probe tip level.
The tympanometric peak in the graph will represent the total
admittance of the ear canal and the middle ear.
Select
Not compensated
to access the field
Scale (Not com-
pensated)
setting.
Otometrics - MADSEN OTOflex 100
147
14 Immittance module settings