B/G numeric com-
ponent com-
pensation for Y
Calculates the middle ear admittance by subtracting the susceptance
(B) and conductance (G) individually from the baseline reference B
and G values (recommended for baseline compensation in 1000 Hz
tympanometry).
Note
•
The curve will never have negative admittance values,
since any difference from a reference point is always pos-
itive. That is why the curve may not look like a traditional tym-
panogram at the negative tail. This is perfectly normal.
This strategy is more important when using high frequency probe
tones (to account for phase shifts during the pressure sweep), but can
also be used for 226 Hz. Phase shifts are much smaller when using
226 Hz and have traditionally been considered negligeable. Hence,
common subtraction of admittance (Y) is typically used for baseline
compensation although it is not as correct.
Curves
Layered curves
Enables a typical ETF-I setting with layered tympanograms.
When
Layered curves
is enabled, this affects a number of areas:
•
If the
Autoscale tympanogram
icon is
selected
, the graph will
be scaled to allow for showing the highest curve. Auto scaling is
still ear dependent.
•
If a 1000 Hz tympanogram is included among the layered curves,
the applicable units and tymp results labels for 1000 Hz probe
tone measurements will also be used for the 226 Hz curves.
•
The widest pressure range used will be applied.
When
Layered curves
is enabled, the curves are denoted by the fol-
lowing graphical conventions.
The selected curve is highlighted and shown as a bold full line.
•
Curve 1, full line
•
Curve 2, dotted line
•
Curve 3, dashed line
Pressure axis
Range
Enables using the same aspect ratio for all pressure sweeps, regard-
less of whether they are for in200 to -400 or +400 to -600.
If the +200 to -400 range is selected, but a wider custom pressure
range is used, the pressure scale will automatically be set to +400 to -
600.
148
Otometrics - MADSEN OTOflex 100
14 Immittance module settings