Ch(ar)g(er) power save
This setting applies when MADSEN OTOflex 100 is placed in a powered charger and uses
rechargeable batteries. After this time in active mode without any user actions on
MADSEN OTOflex 100 or via OTOsuite, MADSEN OTOflex 100 will enter power save
mode where the display changes to a power save screen and the auto start and auto
resume functions stop working. Power save mode reduces power consumption and
allows the batteries to charge.
Note
•
MADSEN OTOflex 100 will only charge while in power save or while
turned off. In active mode, the batteries will still loose power.
Recommended setting: 5 minutes or as short as possible.
Press any key or operate MADSEN OTOflex 100 from OTOsuite to resume full operation.
Ch(ar)g(er) power off
This setting applies when MADSEN OTOflex 100 is placed in a powered charger and uses
rechargeable batteries. After this time period in power save mode, MADSEN OTOflex
100 will power off. Power off allows the batteries to charge and prolongs device life-
time.
Recommended setting: 60 minutes or as short as possible.
No power off before
This setting applies when MADSEN OTOflex 100 is placed in a powered charger and uses
rechargeable batteries.
Prevents automatic power off until a certain hour of the day.
Adjust the setting to the end of your typical working day if you like to keep MADSEN
OTOflex 100 powered and ready for use during working hours.
Set it to 0 to always allow automatic power off.
Localization
Menu item with access to:
•
Language selection
•
Date setting
•
Probe standard
ISO, ANSI, or Ref. EAR-3A.
Set time
Set the year, month, day, hour format, hour and minute settings.
Calib. old warning
For setting up he device to warn when the calibration is one year old. Default is On.
Load factory settings
For loading the preset factory settings.