13.7
ETF-P setup
The following settings are useful:
Initial pressure
Menu
>
The pressure used for the first part of the test.
Default value:
Pos
(itive)
Pos(itive) Pressure
Menu
>
Select the maximum positive pressure to be presented to the ear canal/middle-ear cavity.
100 to 400 daPa, presented in 50 daPa increments.
Default: 400 daPa
Neg(ative) pressure
Menu
>
Select the maximum negative pressure to be presented to the ear canal/middle-ear cav-
ity.
-100 to -600 daPa, presented in 50 daPa increments.
Default: -400 daPa
Pump speed
Menu
>
Changes the speed of the pressure sweep. Indicated in daPa per second.
See
ETF-P (Eustachian Tube Function - Perforated)
.
Note
•
It is highly recommended that the lowest setting (50 daPa/s) is always
used for ETF-P.
Default: 50 daPa
Initial time scale
Menu
>
30-60 seconds. If the test exceeds the time axis, the axis will expand automatically to
show first 40 seconds, then 50 and finally 60 seconds.
Any of the settings available can be predefined for the axis.
Default value: 30 seconds.