ETF-I (Eustachian Tube Function - Intact)
Note
•
This test is available only if MADSEN OTOflex 100 supports diagnostic testing.
Procedure
1. Select the
Tympanometry
test screen.
2. Record a tympanogram.
3. Instruct the patient to perform either Valsalva’s or
Toynbee’s maneuver.
4. Advance to the next curve number.
5. Record a second tympanogram.
6. Compare the tympanograms from step 1 and 3 in the
multilayered tympanogram.
7. It may be useful to repeat the procedure using dif-
ferent techniques and maneuvers in a sequence of test-
ing to fully evaluate the functioning of the Eustachian
tube.
11.4.1
Settings - ETF-I
Recommended settings - ETF-I
Setting
Menu
>
More settings..
>
Auto next curve
The tympanometry graph shows a maximum of 3 curves per ear per patient.
On
:
The curve number is advanced automatically before the tympanogram is recorded (except if the
current curve memory is empty). If you continue measuring, the curve number is cyclically
increased and older tympanograms are overwritten.
Off
:
You must manually toggle the
Curve selector
to the next curve number to preserve a previously
recorded tympanogram.
Default:
On
Menu
>
Layered curves
This view setting is useful when viewing ETF-I results.
Yes
100