11
Testing with the OTOsuite Immittance module
11.1
Sequence testing
A test sequence is a predefined set of automatic tests which can be performed automatically.
MADSEN OTOflex 100 will always perform the most recently selected sequence, regardless of whether it has been selec-
ted in the OTOsuite Immittance module or in MADSEN OTOflex 100.
Note
•
The Reflex Thresholds are included in the test sequences in order to measure Reflex Decay, since the
decay stimulus levels are based on thresholds.
11.1.1
Selecting a test sequence
You can perform the immittance tests in sequences.
–
MADSEN OTOflex 100:
Menu
>
Procedure options..
>
Sequence
–
OTOsuite: In the
Sequence
field in the
Control Panel
click the drop-down list to select.
You can choose between
–
T + RS
(Tympan Reflex Screening)
–
T + RT
(Tympan Reflex Threshold)
–
T + RT + RD
(Tympan Reflex Thr Reflex Decay)
–
Off
(MADSEN OTOflex 100)
The press-and-hold function is disabled, and only the currently selected measurement type will be performed.
11.1.2
Running a test sequence
Starting a sequence
1. Connect to MADSEN OTOflex 100.
2. Select the desired sequence from the drop-down list below the Immittance module Control Panel.
3. Click
Start
to start the selected sequence.
If
Tools > Options > General > Auto start on seal
is set to
Sequence
, the sequence starts as soon as
probe seal is achieved.
Warning
•
If the patient is troubled by the test, stop the test. The test is interrupted and the pump
pressure is relieved immediately.Already measured results are kept.
Pausing or stopping a sequence
If a leak is detected or the probe is removed from the patient's ear, the sequence pauses automatically.
90
Otometrics - MADSEN OTOflex 100
11 Testing with the OTOsuite Immittance module